Model Number 3688 |
Device Problems
No Device Output (1435); Low Battery (2584)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 11/16/2014 |
Event Type
Injury
|
Manufacturer Narrative
|
Results: the observation of inoperable was not confirmed.The observation of battery depletion was confirmed.A review of the ipg register information confirmed the device had declared eri.The device was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.Based on the available information the device met the estimated longevity and had normal battery depletion to the end of life.The estimate longevity was performed referencing the clinicians manual.The ipg was fully functional during analysis.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Manufacturer Narrative
|
Date of explant correct date is (b)(6) 2014.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
It was reported the patient's ipg battery is depleted and as a result is inoperable.The patient underwent surgical intervention to remove and replace the ipg which resolved the issue.
|
|
Manufacturer Narrative
|
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Search Alerts/Recalls
|