• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problem Failure to Advance (2524)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/17/2014
Event Type  Injury  
Event Description
It was reported during the patient's lead revision procedure (reference mfr.Report# 1627487-2014-08473) the physician had trouble placing the lead in the ipg header.The physician tried troubleshooting to no avail.Subsequently, the physician explanted the ipg and replaced it with a new one which has resolved the issue.The patient has effective therapy post-operatively.
 
Manufacturer Narrative
(b)(4).This ipg serial number was included in field advisories.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4310307
MDR Text Key5270398
Report Number1627487-2014-20491
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PP010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 11/17/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2012
Device Model Number3788
Device Lot Number3199437
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/17/2014
Initial Date FDA Received12/09/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/07/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-05242011-002-R
Patient Sequence Number1
Treatment
MODEL 3186, SCS LEAD
Patient Outcome(s) Other;
Patient Age51 YR
-
-