• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problems Computer Software Problem (1112); Pacing Intermittently (1443)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/17/2014
Event Type  malfunction  
Event Description
It was reported and error occurred and pacing intermittent off.The device was returned for repair.No patient complications were reported as a result of this event.
 
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Manufacturer Narrative
Product event summary: analysis confirmed the reported event, the main printed circuit board (pcb) was out of electrical specification.The display wires were not routed properly, they were routed around a display frame screw and were tight once, plugged into the main pcb connector.The keypad and one knob were contaminated, and the encoders were not torqued properly, the torque was out of specification on the low side.
 
Manufacturer Narrative
Further analysis was performed on the main printed circuit board (pcb).Visual inspection revealed no anomalies.Bench analysis revealed no error during power up.Conclusion: the reported event was confirmed by log analysis.An error triggered the out of service error seen by the customer, however analysis was unable to reproduce the error.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4311868
MDR Text Key5255981
Report Number3004593495-2014-00049
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K132924
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 09/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/21/2015
Initial Date FDA Received12/09/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
01/21/2015
Supplement Dates FDA Received02/10/2015
04/10/2015
09/15/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-