Model Number 5392 |
Device Problems
Computer Software Problem (1112); Pacing Intermittently (1443)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/17/2014 |
Event Type
malfunction
|
Event Description
|
It was reported and error occurred and pacing intermittent off.The device was returned for repair.No patient complications were reported as a result of this event.
|
|
Manufacturer Narrative
|
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
|
|
Manufacturer Narrative
|
Product event summary: analysis confirmed the reported event, the main printed circuit board (pcb) was out of electrical specification.The display wires were not routed properly, they were routed around a display frame screw and were tight once, plugged into the main pcb connector.The keypad and one knob were contaminated, and the encoders were not torqued properly, the torque was out of specification on the low side.
|
|
Manufacturer Narrative
|
Further analysis was performed on the main printed circuit board (pcb).Visual inspection revealed no anomalies.Bench analysis revealed no error during power up.Conclusion: the reported event was confirmed by log analysis.An error triggered the out of service error seen by the customer, however analysis was unable to reproduce the error.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|