Model Number 5392
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2014
Event Type  malfunction  
Event Description
It was reported that the external pulse generator had broken black plastic pieces where the ventricular and atrial leads were to be inserted.The generator was returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Event summary: analysis confirmed the customer comment that the output connectors were broken.Analysis also found that the lower case, hanger assembly, keypad, one knob, the main seal and the ring screw were contaminated, that the main world label was damaged and that the encoder nuts torques were out of specification.All found defective parts were replaced.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industr. zone
bayan lepas,my 1190 0
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4313365
MDR Text Key5208647
Report Number3004593495-2014-00055
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132924
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2014
Date Manufacturer Received10/08/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/02/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-