• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMERICAN MEDICAL SYSTEMS (MN) AMS ACTICON NEOSPHINCTER; IMPLANTED FECAL INCONTINENCE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMERICAN MEDICAL SYSTEMS (MN) AMS ACTICON NEOSPHINCTER; IMPLANTED FECAL INCONTINENCE DEVICE Back to Search Results
Catalog Number SEE H10
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Incontinence (1928)
Event Date 11/19/2014
Event Type  Injury  
Event Description
It was reported that the patient had her acticon device replaced due to recurring incontinence.No patient complications were reported in relation with this event.
 
Manufacturer Narrative
Cuff catalog # 72401960; exp.Date:11/12/2013; serial# (b)(4); mfg.Date:11/2008.Balloon catalog #72402106; exp.Date: 03/10/2015; serial (b)(4); mfg.Date: 3/2010.Pump catalog #72402287; exp.Date: 04/12/2015; serial (b)(4); mfg.Date: 4/2010.
 
Manufacturer Narrative
Device evaluation: the explanted device was returned and analyzed.The results indicate the balloon pressure fell below specifications and was related to the reported recurrent incontinence.The pump did not meet specifications, however, it remained functional.The cuff performed within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS ACTICON NEOSPHINCTER
Type of Device
IMPLANTED FECAL INCONTINENCE DEVICE
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS (MN)
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS (MN)
10700 bren road w
minnetonka MN 55343
Manufacturer Contact
sharon zurn
10700 bren road w
minnetonka, MN 55343
9529306000
MDR Report Key4316599
MDR Text Key5211374
Report Number2183959-2014-00546
Device Sequence Number1
Product Code MIP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue NumberSEE H10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/25/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age42 YR
-
-