Device Problems
Break (1069); Material Integrity Problem (2978)
Patient Problem
No Known Impact Or Consequence To Patient (2692)
Event Date 03/25/2013
Event Type
malfunction
Event Description
Customer called in and stated that the head deck bolts from their bed broke.
Manufacturer Narrative
A review of complaints took place and the firm has concluded that this complaint should have been filed as a mdr.The bed frame had the right and left head deck pivot point bolt shear causing the head section of the bed to detach from the mid-section of the bed frame.During the review of complaints there were similar occurrences that the firm found and have filed mdrs for these complaints as well.The firm is monitoring for any future complaints.