The results of the investigation are inconclusive since the device was not returned for analysis; however, photos of the explanted valve were received from the field.There appeared to be dense granular tissue formation on portions of both the inflow and outflow surfaces of each cusp, and along each stent post.There appeared to also be white tissue formation on the visible sections of the sewing cuff, a tear or cut at the free edge of one of the cusps, and a tear or cut at the base of another cusp along the stent post.Due to the limited view of the device, no additional observations were able to be made.The device history record was reviewed, which showed that each manufacturing and inspection operation was performed and indicated complete in accordance with sjm specifications and procedures.Based on the information received, the cause of the reported incident could not be conclusively determined.
|