The customer states that on (b)(6) 2014 after the dialysis treatment was completed, the nurse discovered that the catheter the cuff section of the catheter was now visible outside of the skin.The catheter had originally been implanted on (b)(6) 2014.On (b)(6) 2014, the catheter was removed and replaced with a new catheter.The patient was hospitalized; and no information was provided related to any required treatment.No sample was returned for further evaluation.A device history record (dhr) review was performed for the lot number identified by the customer, 322716x.The dhr review indicated that there was no quality issues associated with the reported defect mode.The complaint was investigated, but since no sample was returned to covidien the defect could not be confirmed and the probable root cause cannot be determined.Complaint trending information is being reviewed on a monthly basis by covidien and if a trend is observed, actions are taken as necessary.It should be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling) are in place to prevent nonconforming product from leaving the manufacturing operations.This information will be utilized for tracking and trending purposes to determine the need for corrective action.
|