• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO TCM II COOLING AND HEATING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO TCM II COOLING AND HEATING SYSTEM Back to Search Results
Model Number 4415
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/17/2014
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the cooler heater unit was leaking in the front by the motor.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The user facility's biomed engineer (biomed) brought the machine down to shop and found it was leaking from the front.He attempted to order parts but was informed he was not trained.The field service rep (fsr) performed preventive maintenance (pm) on 09/09/2014.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO TCM II COOLING AND HEATING SYSTEM
Type of Device
TCM II COOLING AND HEATING SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4340547
MDR Text Key5172339
Report Number1828100-2014-01045
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/18/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-