The customer states that the catheter was implanted in the patient on (b)(6) 2014.Per the complaint notes, the incident date was (b)(6) 2014.The nurse discovered that the catheter had become loose and the cuff section of the catheter was now visible outside of the skin.The catheter was removed and medical intervention was required.No sample was returned for further evaluation.A device history record (dhr) review was performed for the lot number identified by the customer, 322716x.The complaint was investigated, but since no sample was returned to covidien the defect could not be confirmed and the probable root cause cannot be determined.Complaint trending information is being reviewed on a monthly basis by covidien and if a trend is observed, actions are taken as necessary.It should be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling) are in place to prevent nonconforming product from leaving the manufacturing operations.This information will be utilized for tracking and trending purposes to determine the need for corrective action.
|