Model Number 802100
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2014
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the central control monitor (ccm) screen locked up.User attempted to select a perfusion configuration and it locked up.The device was not changed out, as they are still using the unit, they just reboot and the ccm is fine.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Evaluation is in progress, but not yet concluded.The field service representative (fsr) replaced the central control monitor (ccm) and configured the new ccm.The unit operated to manufacturer specifications and was returned to clinical use.The suspect ccm was returned to the manufacturer for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
APS 1 (CCM)
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4350914
MDR Text Key21770925
Report Number1828100-2014-01067
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number802100
Device Catalogue Number802100
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/03/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/21/2014
Initial Date FDA Received12/15/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-