Catalog Number 595000-001
Device Problem Filling Problem (1233)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/20/2014
Event Type  malfunction  
Event Description
The customer reported that the freedom driver exhibited fluctuating fill volumes while supporting a patient.The customer also reported that the patient was asymptomatic.The patient was switched to the backup freedom driver.There was no reported adverse patient impact.This alleged failure mode poses a low risk to the patient because it does not prevent the freedom driver from performing its life-sustaining functions.Although the patient experienced fluctuating fill volumes, the freedom driver continued to function as intended.An investigation will be conducted by syncardia and the results will be provided in a supplemental mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
CIRCULATORY ASSIST DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
tucson AZ
Manufacturer Contact
carole marcot, esq., vp
1992 e. silverlake rd.
tucson, AZ 85713
5205451234
MDR Report Key4352398
MDR Text Key19309282
Report Number3003761017-2014-00285
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number595000-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/20/2014
Initial Date FDA Received12/11/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age62 YR
-
-