MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1074300-23
Device Problems Detachment Of Device Component (1104); Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/02/2014
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat a 99% stenosed lesion in the mid to distal right coronary artery with mild calcification.Pre-dilatation was performed.A 3.0 x 15 mm xience xpedition stent was implanted distally; however, a dissection occurred.The 3.0 x 23 mm xience xpedition stent delivery system (sds) was advanced to treat the dissection; however, the sds failed to cross and the hypotube separated outside the anatomy.A new xience xpedition sds was used to treat the dissection successfully.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Concomitant products: stent: 3.0 x 15 mm xience xpedition; guide wire: route; guide cath: brite tip 6f jr4.0.The 3.0 x 15 mm xience xpedition referenced is being filed under a separate medwatch report.Evaluation summary: the device was returned for analysis.The separation was able to be confirmed.The failure to advance could not be replicated in a testing environment as it was based on operational circumstances.Based on a visual and dimensional inspection of the returned device, there is no indication of a product deficiency.A review of the lot history record revealed no non-conformances that would have contributed to the reported event.A query/review of the complaint history of the reported lot did not indicate a manufacturing issue.Based on the reviewed information, no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key4353858
MDR Text Key18293614
Report Number2024168-2014-08412
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 12/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/04/2016
Device Catalogue Number1074300-23
Device Lot Number4052941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/03/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CONCOMITANT MEDICAL DEVICES
-
-