• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. DUAL COOLER/HEATER; SARNS DUAL COOLER/HEATER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. DUAL COOLER/HEATER; SARNS DUAL COOLER/HEATER Back to Search Results
Model Number 11160
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Patient Involvement (2645)
Event Date 11/25/2014
Event Type  malfunction  
Event Description
The user facility's biomedical engineer (biomed) reported that during preventive maintenance (pm) of the device, the cooler heater unit went into "overtemp" at 42 degrees celsius.There was no patient involvement.
 
Manufacturer Narrative
This customer is not trained by the manufacturer to perform repairs on this unit.The field service representatives (fsr) spoke with the biomed.This was user error per the biomed.The perfusionist (ccp) using the unit was not familiar with it and was not aware that the unit was not familiar with it and was not aware that the unit will go "overtemp" for a second and then come back down.They are canceling the service ticket for this unit.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL COOLER/HEATER
Type of Device
SARNS DUAL COOLER/HEATER
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4355836
MDR Text Key5141270
Report Number1828100-2014-01098
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PREAMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number11160
Device Catalogue Number11160
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/25/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-