Catalog Number 1009529-18
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Intimal Dissection (1333)
Event Date 12/01/2014
Event Type  Injury  
Event Description
It was reported that during a procedure of the moderately tortuous, mildly calcified, 99% stenosed, mid right coronary artery (rca) pre-dilatation was completed using a 1.5 x 12 mm unspecified balloon dilatation catheter (bdc) and a 3.0 x 18 mm xience v stent was implanted in the lesion.A distal edge dissection was noted.Another xience v stent was used as treatment.There was no reported adverse patient sequela.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The stent remains in the vessel.It is indicated that the device is not returning for evaluation; therefore, a failure analysis of the complaint device could not be completed.A review of the lot history record revealed no non-conformances that would have contributed to the reported event.Dissection is listed in the xience v everolimus eluting coronary stent systems instructions for use (ifu) as a known patient effect of coronary stenting procedures.Although a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined, there is no indication of a product quality deficiency with respect to manufacturing, design or labeling.Based on the information reviewed, there is no evidence to indicate the presence of a product deficiency.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key4358268
MDR Text Key17587181
Report Number2024168-2014-08466
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P070015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 12/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/19/2016
Device Catalogue Number1009529-18
Device Lot Number4051541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2014
Initial Date FDA Received12/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BALLOON DILATATION CATHETER: 1.5X12MM
Patient Outcome(s) Required Intervention;
Patient Age52 YR
Patient Weight63
-
-