• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310R
Device Problems Degraded (1153); Volume Accuracy Problem (1675)
Patient Problem No Patient Involvement (2645)
Event Date 12/08/2014
Event Type  malfunction  
Event Description
During evaluation of a returned homechoice device, a baxter technician determined the device failed fluid volume accuracy testing.No additional information is available.
 
Manufacturer Narrative
Complaint no: cmplnt-(b)(4).The homechoice (hc) device was returned and evaluated.A service history review was performed and revealed no issues that could have caused or contributed to the rite (returned instrument test evaluation) failure.The external/internal inspection was performed and the device passed, along with all of the electrical testing.During rite functional testing the volumetric accuracy test failed.A detailed inspection of the door assembly showed that the piston foam was deteriorated.The cause of the volumetric accuracy failure was determined to be the deteriorated piston foam.The piston foam was to be scrapped and the device was sent for servicing.Should additional relevant information become available, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4358993
MDR Text Key5119814
Report Number1416980-2014-47325
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C8310R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2014
Date Manufacturer Received12/08/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-