Model Number 3207-36
Device Problem Failure to Interrogate (1332)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/13/2014
Event Type  Injury  
Manufacturer Narrative
Evaluation description: the reported inability to interrogate was confirmed in the laboratory and was due to premature battery depletion.Review of the information provided to technical services indicated that the longevity was below expected limits.The device was tested on the bench and no anomalies were found.The original battery was returned to the vendor for further evaluation and no anomaly was found.The cause of the premature battery depletion could not be determined.
 
Event Description
It was reported that upon interrogation, the device could not be interrogated.Patient was asymptomatic.Device was explanted and replaced.The patient was fine post procedure.
 
Manufacturer Narrative
Device evaluation anticipated, but not yet begun.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE RF CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sharon mathew
645 almanor avenue
sunnyvale, CA 94085
4085226327
MDR Report Key4359239
MDR Text Key13076625
Report Number2938836-2014-19151
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2010
Device Model Number3207-36
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/13/2014
Initial Date FDA Received12/23/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/22/2015
03/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age90 YR
-
-