• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC., CRMD QUADRA ASSURA CRT-D, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC., CRMD QUADRA ASSURA CRT-D, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3365-40Q
Device Problem Under-Sensing (1661)
Patient Problems Cardiomyopathy (1764); Death (1802)
Event Date 11/24/2014
Event Type  Death  
Event Description
It was reported that the patient expired.Review of strips revealed undersensing of small r waves.Although atp therapy was delivered, there were no high voltage therapies.It was noted that the patient had severe end-stage cardiomyopathy and suspected electrolyte imbalance.No autopsy was performed.There is no known allegation from a health professional that the death was related to the device.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D, DF-4 CONNECTOR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sharon mathew
645 almanor avenue
sunnyvale, CA 94085
4085226327
MDR Report Key4359282
MDR Text Key5236090
Report Number2938836-2014-19143
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2016
Device Model NumberCD3365-40Q
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/25/2014
Initial Date FDA Received12/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/23/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age81 YR
-
-