SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM A/C POWER SUPPLY Back to Search Results
Catalog Number 295050-001
Device Problems Connection Problem (2900); Power Problem (3010)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 11/21/2014
Event Type  malfunction  
Event Description
The freedom a/c power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that the plug on a pt's freedom a/c power supply would not connect to the freedom power adaptor.The pt was provided with a replacement a/c power supply.There was no reported adverse pt impact.This alleged failure mode poses a low risk to the pt because it did not prevent the freedom driver from performing its life-sustaining functions.The freedom driver has a redundant power source of onboard batteries.An investigation will be conducted by syncardia.The results of the investigation will be provided in a supplemental mdr.
 
Manufacturer Narrative
The ac power supply cable was disassembled, the connector key slot was aligned with the connector cover key tab and the correct mating configuration was obtained.After the ac power supply cable was reassembled, the ac power supply passed all final performance testing.This failure mode poses a low risk to a pt because the freedom ac power supply was not in use by a pt at the time of the customer-reported issue.Pts are provided with two freedom ac power supplies.In addition, it would not prevent a freedom driver from performing its life-sustaining functions.The freedom driver is equipped with redundant power sources, including multiple rechargeable freedom onboard batteries and a car charger.The freedom ac power supply met all required acceptance criteria and the unit will be returned to finished goods.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
 
Event Description
The ac power supply was returned to syncardia for evaluation.Visual inspection of the freedom ac power supply revealed that the cable connector cover was installed backwards.This is the root cause of the customer-reported inability to connect the ac power supply to a freedom power adapter.It is likely that the cable connector cover was installed incorrectly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM A/C POWER SUPPLY
Type of Device
POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
tucson AZ
Manufacturer Contact
carole marcot, esq, vp
1992 e. silverlake rd.
tucson, AZ 85713
5205451234
MDR Report Key4359950
MDR Text Key5112294
Report Number3003761017-2014-00288
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,health professional,use
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 11/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number295050-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/24/2014
Initial Date FDA Received12/15/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/11/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-