The ac power supply cable was disassembled, the connector key slot was aligned with the connector cover key tab and the correct mating configuration was obtained.After the ac power supply cable was reassembled, the ac power supply passed all final performance testing.This failure mode poses a low risk to a pt because the freedom ac power supply was not in use by a pt at the time of the customer-reported issue.Pts are provided with two freedom ac power supplies.In addition, it would not prevent a freedom driver from performing its life-sustaining functions.The freedom driver is equipped with redundant power sources, including multiple rechargeable freedom onboard batteries and a car charger.The freedom ac power supply met all required acceptance criteria and the unit will be returned to finished goods.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
|