The freedom ac power supply was not in use by a patient.The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.A syncardia clinical support specialist reported that the connector on the ac power supply would not fit into the power adaptor.Visual inspection of the freedom ac power supply revealed the output power connector terminals were misaligned.This is the root cause of the customer reported issue, which prevented the ac power supply from connecting well with the freedom power adaptor.The cause of the misaligned connector terminals is unknown; however, it is likely that this occured in the manufacturing process.This failure mode poses a low risk to a patient because the freedom ac power supply was not in use by a patient at the time of the customer reported issue.In addition, it would not prevent a freedom driver from performing its life-sustaining functions.The freedom driver is equipped with redundant power sources, including multiple rechargeable freedom onboard batteries and a car charger.The freedom ac power supply was taken out of service.
|