• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM DRIVER; CIRCULATORY ASSIST DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM DRIVER; CIRCULATORY ASSIST DEVICE Back to Search Results
Catalog Number 595000-001
Device Problems Break (1069); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/13/2014
Event Type  malfunction  
Event Description
The customer reported that the freedom power adaptor could not be connected to the freedom driver because the connector receptacle on the driver that connects the power adaptor to the driver was broken.The pt was switched to the backup driver.There was no reported adverse pt impact.Although the freedom power adaptor could not be connected to the freedom driver, the driver continued to function as intended.This alleged failure mode poses a low risk to the pt because it does not prevent the freedom driver from performing its life-sustaining functions.The freedom driver has a redundant power source of onboard batteries.The freedom driver will be returned to syncardia for eval.The results of the investigation will be provided in a supplemental mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
CIRCULATORY ASSIST DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
tucson AZ
Manufacturer (Section G)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake rd.
tucson AZ 85713
Manufacturer Contact
carole marcot, esq, vp regulator
1992 e. silverlake rd.
ce 2474
tucson, AZ 85713
5205451234
MDR Report Key4360761
MDR Text Key5119859
Report Number3003761017-2014-00274
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date 11/17/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number595000-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2014
Initial Date FDA Received12/05/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age45 YR
-
-