MAUDE Adverse Event Report: BAXTER HEALTHCARE - IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1008KP
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 11/07/2014
Event Type  malfunction  
Event Description
It was reported that a large volume infusor leaked from an unknown source into the housing of the device.This occurred during filling.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Complaint no: cmplnt-(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the sample was received for evaluation.Visual inspection noted fluid inside the housing which indicated leakage had occurred.The reported condition was verified.The cause of the condition was determined to be a missing film wrap component; however, the cause of the missing component was not determined.A capa has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4363019
MDR Text Key5260269
Report Number1416980-2014-47526
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K071222
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 12/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/01/2017
Device Catalogue Number2C1008KP
Device Lot Number14D033
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/02/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/19/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-