It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the level sensor was falsely alerting and stopping the roller pump.The device was not changed out, as the perfusionist (ccp) shut off the alert sensor and observed the cardiotomy herself to finish case.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per the clinical review on (b)(6) 2014: during cpb, a low level alert condition occurred and the arterial pump paused (went to 0 speed) and returned to previous speed when the fluid level returned to previous speed when the fluid level returned above the sensor.This caught the ccp off guard, as she was expecting a "message only" pump response with a low level alert.The ccp did not know how to re-configure the alert to "message only", so she elected to turn the level sensor off and not use it for the remainder of the procedure.She visually monitored the reservoir blood level for the remainder of the procedure.The case was completed successfully, without delay and without associated blood loss.There was no harm observed.
|