Device 1 of 2.Reference mfr.Report#: 1627487-2014-23839.The patient reported he experienced heating at the ipg pocket site while charging.Subsequently, he underwent surgical intervention in (b)(6) 2011 where his scs system was explanted which resolved the reported issue.Please note: the exact event date is unknown at this time.On 08/01/2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients. an increase in prior non-reported heating while charging events and other non-reported events was expected.
|