MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION LAMITRODE TRIPOLE 16C LEAD, 60 CM; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION LAMITRODE TRIPOLE 16C LEAD, 60 CM; SCS LEAD Back to Search Results
Model Number 3214
Device Problems Therapy Delivered to Incorrect Body Area (1508); Device Operates Differently Than Expected (2913)
Patient Problems Therapeutic Effects, Unexpected (2099); Inadequate Pain Relief (2388)
Event Date 12/11/2014
Event Type  Injury  
Event Description
It was reported the patient is experiencing stimulation in the incorrect location.Reprogramming was unable to resolve the issue.X-rays were taken and indicated the lead had migrated.Surgical intervention may take place at a later date to address the issue.
 
Event Description
Follow-up information indicated surgical intervention was undertaken on (b)(6)2014.The physician relocated the lead and the patient reported effective stimulation coverage postoperatively.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMITRODE TRIPOLE 16C LEAD, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4370068
MDR Text Key5115774
Report Number1627487-2014-15925
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/30/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/28/2016
Device Model Number3214
Device Lot Number4610259
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/30/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3789, SCS IPG
Patient Outcome(s) Other;
Patient Age58 YR
-
-