Model Number 3608 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
Device Overstimulation of Tissue (1991)
|
Event Type
Injury
|
Event Description
|
It was reported the patient's (b)(6) ipg is no longer communicating after turning her scs system off for a few days due to experiencing overstimulation approximately two months ago.In turn, the patient underwent surgical intervention to explant and replace the ipg which resolved the issue.
|
|
Manufacturer Narrative
|
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Manufacturer Narrative
|
Results: a visual inspection of the ipg did not reveal any discrepancies.Communication with the device as received was unsuccessful when using a patient programmer and wdl utility.Electrical testing of the ipg confirmed the ipg battery voltage was below the minimum to sustain communication and stimulation.The device was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.The auto test confirmed the integrity of the header assembly to ensure there were no issues that could have contributed to the patient feeling a shocking sensation.The observation of the ipg not working was due to the ipg battery having a low voltage.The root cause of the low battery voltage was not ascertained.An accurate longevity of the ipg could not be performed as the programming history was unknown.The observation of shocking was not confirmed through product analysis.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Search Alerts/Recalls
|