Model Number 3716 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
Burning Sensation (2146)
|
Event Date 12/10/2014 |
Event Type
Injury
|
Event Description
|
Follow up information identified the patient is not experiencing discomfort.The patient has always experienced warmth with both the charging systems.
|
|
Manufacturer Narrative
|
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
Device #1 of 2.Reference mfr.Report: 1627487-2014-22057.It was reported the patient was undergoing surgical intervention to replace her ipg (reference mfr.Report: 1627487-2014-08523).Pre-operatively the patient reported she experienced warmth while recharging.In addition, it is unknown if she was also experiencing discomfort.On 08/01/2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients. an increase in prior non-reported heating while charging events and other non-reported events was expected.
|
|
Manufacturer Narrative
|
This ipg serial number was included in a field correction.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|