• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC., CRMD QUADRA ASSURA MP CRT-D, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC., CRMD QUADRA ASSURA MP CRT-D, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3371-40Q
Device Problems Loose or Intermittent Connection (1371); Over-Sensing (1438)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/31/2014
Event Type  Injury  
Event Description
It was reported that non sustained ventricular tachycardia due to myopotential noise was observed.The device was reprogrammed.A loose connection was suspected and the pocket was revised.The lead was unscrewed, cleaned and reconnected to the header.Post revision, noise was not seen with pocket manipulation.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP CRT-D, DF-4 CONNECTOR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sharon mathew
645 almanor avenue
sunnyvale, CA 94085
4085226327
MDR Report Key4372188
MDR Text Key5113688
Report Number2938836-2014-19310
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2016
Device Model NumberCD3371-40Q
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/31/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/20/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-