ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Increased Sensitivity (2065); Burning Sensation (2146)
Event Date 12/10/2014
Event Type  Injury  
Manufacturer Narrative
Result: the returned ipg communicated with all lab utilities, passed functional testing and when programmed, did not display any anomalous outputs when monitored on an oscilloscope.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient experienced burning sensation at the ipg site regardless of stimulation and charging.Sjm representative confirmed the ipg being superficial and the ipg site being tender.As a result, the physician explanted the scs ipg and replaced it and also the physician enlarged the pocket to place the new ipg deeper.
 
Manufacturer Narrative
Udi (di): (b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4375862
MDR Text Key12977653
Report Number1627487-2014-20528
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/28/2015
Device Model Number3788
Device Lot Number3932311
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/24/2015
Initial Date FDA Received12/31/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/16/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3386(2), SCS EXTENSION; MODEL 3186(2), SCS LEAD
Patient Outcome(s) Other;
Patient Age63 YR
-
-