PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) Back to Search Results
Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 11/07/2014
Event Type  malfunction  
Event Description
The customer contacted physio-control to report that a service indicator was displayed on their device.The customer was advised to replace the coin cell battery in the device.After the customer replaced the coin battery, they observed the device reset itself during boot up.There was no patient use associated with the reported event.
 
Manufacturer Narrative
(b)(4).Physio-control evaluated the device and verified the reported failure.Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Manufacturer Narrative
Physio-control replaced the system controller and user interface pcb assemblies and the flex cable assembly which connects the user interface pcb with the pcb stack.Proper device was then observed through functional and performance testing and the device was returned to the customer for use.Physio-control determined that the flex cable which connects the user interface pcb assembly to the pcb stack was the cause of the reported failure.
 
Manufacturer Narrative
Further testing of the removed flex cable assembly revealed that there was damage to the flex cable near the p21 plug connector which led to the reported boot failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4377421
MDR Text Key5289185
Report Number3015876-2015-00001
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 12/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2014
Is the Reporter a Health Professional? Yes
Device Age9 YR
Event Location Hospital
Date Manufacturer Received01/27/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/07/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-