• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON C 16-CHANNEL IPG; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON C 16-CHANNEL IPG; SCS IPG Back to Search Results
Model Number 3688
Device Problems No Device Output (1435); Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896)
Patient Problem Inadequate Pain Relief (2388)
Event Date 12/05/2014
Event Type  Injury  
Event Description
Follow up information identified the patient underwent surgical intervention to remove and replace her ipg which resolved the issue.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient's programmer displayed a communication error code.Troubleshooting was unable to resolve the issue.A replacement programmer was sent to the patient and did not resolve the issue.Surgical intervention may be pending to address this issue.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON C 16-CHANNEL IPG
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL NEUROMODULATION DIVISION
6901 preston road
plano TX 75024
MDR Report Key4377824
MDR Text Key5400410
Report Number1627487-2015-26002
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2014
Device Model Number3688
Device Lot Number3598569
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/12/2015
Initial Date FDA Received01/02/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/02/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/23/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3219, SCS LEAD
Patient Outcome(s) Other;
Patient Age77 YR
-
-