MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Patient involvement (2648)
Event Date 10/30/2014
Event Type  Other  
Event Description
Sorin group (b)(4) received a report that the touch screen of the s5 roller pump was unresponsive during set up.There was no patient involvement.
 
Manufacturer Narrative
There was no patient involvement.Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the touch screen of the s5 roller pump was unresponsive during set up.There was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Manufacturer Narrative
(b)(4).Sorin group (b)(4) received a report that the touch screen of the s5 roller pump was unresponsive during set up.There was no patient involvement.A sorin group field service representative was dispatched to the facility to investigate.While at the facility, the service representative checked the can bus device history for touch screen errors and confirmed the issue.The gasket and touch screen were replaced and the unit was functionally tested.No further issues were discovered.A review of the dhr could not identify any concessions, deviations or non-conformities relevant to the reported failure.This is a known issue and a capa has been opened to address it.Change order (b)(4) has already been implemented as a corrective action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich D-809 39
GM  D-80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key4385962
MDR Text Key20018440
Report Number1718850-2014-00458
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Unknown
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/12/2014
Initial Date FDA Received12/11/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/29/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-