Model Number 3200
Device Problem Device Sensing Problem (2917)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2014
Event Type  malfunction  
Event Description
Hill-rom received a report from a hill-rom tech stating that the bed exit would not send signal to nurse's station.The bed was located at the account in medsurg.There was no pt/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
 
Manufacturer Narrative
The hill-rom tech found the communication cable inoperable.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unk if the facility performs preventative maintenance on their beds.The tech replaced the communication cable to resolve the issue.Based on this info, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
8129312869
MDR Report Key4390527
MDR Text Key5431318
Report Number1824206-2014-02910
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/04/2014
Initial Date FDA Received12/30/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-