Livanova (b)(4) manufactures the s5 cardioplegia sensor module.The incident occurred in mineola, new york.This medwatch report is being filed on behalf of livanova (b)(4).The livanova field service representative was dispatched to the facility to investigate.The service representative was able to reproduce the issue and observed that the cardioplegia pressure was reading low.The cardioplegia sensor module was replaced and subsequent functional verification testing was completed without further issues.The device was returned to service.The replaced module was returned to livanova (b)(4) for further investigation.The root cause of the described error was confirmed to be a result of poor welding.Two of the pins were found to be welded together.The welding was removed and the module was tested for a 24 hour test run under p-can control with different settings.No further deviations occurred.For saftey reasons, the module was scrapped.A review of the dhr identified that the module failed several automatic test cases that were performed during manufacturing.Upon manual testing, one of the test cases passed when it should have failed.This is the first report for this type of issue.
|