CPI - DEL CARIBE ENDOTAK RELIANCE Back to Search Results
Model Number 0292
Device Problems Failure to Capture (1081); Pocket Stimulation (1463); Device Dislodged or Dislocated (2923)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/20/2014
Event Type  Injury  
Event Description
Boston scientific received information that this right ventricular (rv) lead exhibited high thresholds, muscle stimulation and loss of capture due to dislodgement.The lead was successfully repositioned.No additional adverse patient effects were reported.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
(b)(4).As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 55112
6515824786
MDR Report Key4425018
MDR Text Key5165853
Report Number2124215-2015-00782
Device Sequence Number1
Product Code NVY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/16/2016
Device Model Number0292
Other Device ID NumberRELIANCE 4-SITE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received11/21/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/16/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0292; E162; 4135
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age68 YR
-
-