Model Number TF-23A
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Aortic Insufficiency (1715); Heart Failure (2206)
Event Date 12/23/2014
Event Type  Injury  
Event Description
A 23mm sjm trifecta aortic tissue valve was implanted (b)(6) 2012.In december 2014, the 23mm tissue valve was explanted due to heart failure and aortic insufficiency.At that time, the tee suggested a dehisced partial leaflet.At explant, one of the cusps was found detached from the stent base and prolapsed.After excising the native valve area and debriding annular scar tissue, a 21mm edwards magna valve was implanted.The patient was in stable condition post-operatively.
 
Manufacturer Narrative
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.A review of the device history record was not possible since the serial number was unavailable.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM TRIFECTA VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA)
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 3131 0-26
BR  31310-260
Manufacturer (Section G)
ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA)
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 3131 0-26
BR   31310-260
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key4428552
MDR Text Key5371860
Report Number3001883144-2015-00005
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberTF-23A
Device Catalogue NumberTF-23A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/26/2014
Initial Date FDA Received01/16/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
-
-