The following description of event was documented by a carefusion tech support specialist in response to an email from the distributor in (b)(6)."obf pn 03920a microblender getting 27 set at 21, getting 94 set at 100.Blender failed pre-delivery inspection.Blender is reading high on 21% o2 values, and low on 100% o2 values.Unable to resolve through calibration.".
|
The foreign distributor did not submit a user facility/importer report to the manufacturer.(b)(4).The alleged faulty device was received by carefusion on (b)(4) 2015, routed to the carefusion failure analysis lab and staged for eval.Once the eval is complete, a f/u medwatch report will be submitted.
|
Failure analysis: microblender pn: 03920a, sn: (b)(4) was received into the carefusion failure analysis lab for investigation.The carefusion failure analysis lab technician examined the microblender and found that it is out of calibration.The microblender was tested on test station # 10464 per drs (b)(4) and found that it is out of calibration at 21%, 30%, 60%, 90% and 100% o2.At end point calibration sequence # 1, 21%, spec.Is 20.9 to 22.9% reads 24.3%.At set point calibration sequence # 2, 30%, spec.Is 28.0 to 32.0% reads 24.4%.At set point calibration sequence # 3, 60%, spec.Is 58.0 to 62.0% reads 55.2%.At set point calibration sequence # 4, 90%, spec.Is 88.0 to 92.0% reads 89.1%.At set point calibration sequence # 5, 100%, spec.Is 98.0 to 100.0% reads 97.0%.At set point calibration sequence # 7, 60%, spec.Is 58.0 to 62.0% reads 55.2%.At set point calibration sequence # 8, 60%, spec.Is 57.0 to 63.0% reads 55.7%.At set point calibration sequence # 9, 60%, spec.Is 57.0 to 63.0% reads 54.9%.The microblender could not be recalibrated by adjusting the control knob and or valve seat.The poppet pn: 03929 was replaced and was then, able to recalibrate the microblender.Finding/root-cause: duplicated, the microblender is out of calibration at 21% and 100%, complaint allegation.Also, found that unit is out of calibration at 30%, 60% and 90%.The microblender has a defective poppet pn: 03929.
|