• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Fracture (1260); Device Operates Differently Than Expected (2913); Impedance Problem (2950)
Patient Problem No Code Available (3191)
Event Date 12/16/2014
Event Type  Injury  
Event Description
The patient has 2 leads (from the same lot) implanted as part of her scs system.It was reported the patient is without stimulation.Diagnostic testing revealed invalid impedance readings.X-rays showed the patient's right lead was fractured.The patient may undergo surgical intervention as the next course of action.
 
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Follow-up information revealed the patient underwent surgical intervention on (b)(4) 2015, where the patient's lead was explanted and replaced.During lead revision, the physician decided to electively explant and replace the patient's existing ipg with a new model.Effective stimulation was recaptured postoperative.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano, texas
Manufacturer Contact
st. jude medical - neuromodulation
6901 preston rd
plano 75040
MDR Report Key4439068
MDR Text Key5379654
Report Number1627487-2015-23040
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2015
Device Model Number3186
Device Lot Number3905812
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/09/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/09/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3788, SCS IPG, IMPLANT DATE; MODEL 1192 (2), SCS ANCHOR, IMPLANT DATE
Patient Outcome(s) Other;
Patient Age54 YR
-
-