• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS BATT F/532.001+532.010 12V STD; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS BATT F/532.001+532.010 12V STD; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.003
Device Problem Device Difficult to Setup or Prepare (1487)
Patient Problem No Information (3190)
Event Type  malfunction  
Event Description
It was reported from (b)(6) that battery device did not load.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device was worn out.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If information is obtained that was not available, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATT F/532.001+532.010 12V STD
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4443256
MDR Text Key5387491
Report Number3009450871-2015-10100
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,company representati
Reporter Occupation Other
Type of Report Initial
Report Date 01/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.003
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received08/04/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/29/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-