Customer reported to horiba medical (horiba) that the compressor connected to their abx pentra 400 chemistry analyzer is reporting compressor "errors" with blown fuses.There was no injury or illness in connection with this incident.A horiba medical field service representative (fsr) was dispatched to determine if the instrument was malfunctioning.The fsr confirmed that the cooling unit power receptacle melted on the inside where fuse burned out with contacts missing on the other side of fuse box.The fsr replaced the power receptacle and fuses, and verified that the instrument settings were correct and that the analyzer was functioning.The damaged power cable connectors will be returned to horiba irvine for evaluation.
|