• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM Back to Search Results
Catalog Number 1012580-16
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/25/2014
Event Type  Injury  
Event Description
It was reported that the procedure was to treat a mid left anterior descending artery.A 2.8 x 16 mm graftmaster was implanted to treat a perforation caused by another device; however, it did not completely seal the perforation and the patient was sent to surgery for coronary artery bypass graft (cabg).The patient is doing fine.No additional information was provided.
 
Manufacturer Narrative
(b)(4).It is indicated that the device is not returning for evaluation; therefore, a failure analysis of the complaint device could not be completed.A review of the lot history record revealed no non-conformances.A query of the complaint handling database revealed no other similar incidents reported from this lot.Based on the information reviewed, there is no indication of a product deficiency.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Type of Device
CORONARY STENT GRAFT
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key4446552
MDR Text Key5313245
Report Number2024168-2015-00369
Device Sequence Number1
Product Code MAF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H000001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 01/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2016
Device Catalogue Number1012580-16
Device Lot Number4061641
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/06/2015
Initial Date FDA Received01/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age67 YR
-
-