Device Problem Extrusion (2934)
Patient Problems Adhesion(s) (1695); Diarrhea (1811); Unspecified Infection (1930); Pain (1994)
Event Date 06/27/2014
Event Type  Injury  
Event Description
It was reported the patient had his action removed due to infection, diarrhea, exposure, and anal pain.The patient had experienced diarrhea and anal pain; anterior exposure and positive blood cultures were discovered during examination.The action was removed due to infection from exposure.Abdominal adhesions were also discovered.No additional patient complications were reported in relation to this event.
 
Manufacturer Narrative
Additional information: cuff: catalog #: 72401982, expiration date: 4/02/2018, serial #: (b)(4), manufacture date: 4/2013.Pump: catalog #: 72402287, expiration date: 10/09/2014, serial #: (b)(4), manufacture date: 10/2013.Balloon: catalog #: 72402106, expiration date: 04/02/2018, serial #: (b)(4), manufacture date: 04/2013.The explanted action device was visually inspected and functionally tested.The pump performed within specifications.There was a leak in the cuff that was the result of a sharp instrument which probably occurred during removal.The balloon tested at 94.1 cmh2o.It is rated at 101-110 cmh2o.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS ACTICON NEOSPHINCTER
Type of Device
IMPLANTED FECAL INCONTINENCE DEVICE
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS (MN)
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS (MN)
10700 bren road w
minnetonka MN 55343
Manufacturer Contact
sharon zurn
10700 bren road w
minnetonka, MN 55343
9529306000
MDR Report Key4454400
MDR Text Key5423504
Report Number2183959-2015-00044
Device Sequence Number1
Product Code MIP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/16/2015
Initial Date FDA Received01/27/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age59 YR
-
-