Model Number 3200
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/02/2015
Event Type  malfunction  
Event Description
Hill-rom received a report from the account stating the head section would not lower when left cpr lever was used.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Manufacturer Narrative
The hill-rom technician found the bracket was bent and the pull handle was broken.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2013.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the bracket, pull handle and labeling to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM INC.
batesville IN
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key4471076
MDR Text Key22160629
Report Number1824206-2015-00135
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Unknown
Type of Report Initial
Report Date 01/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/02/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-