Model #: mcl-s22.Device available for evaluation? yes returned to manufacturer: january 29, 2015.Date received by manufacturer: february 2, 2015 h3: product evaluation: analysis for mcl-s21, lot 090104mf3, found that the active part of the instrument is broken.The pin between the jaws did not resist.The tube is bent and there are impacts on the jaws.All the parts are still on the instrument.There is no risk of fragment in the patient.No material or manufacturing defect was found.The most probable cause of this breakage is an excessive effort during the use or the reprocessing of the instrument.Analysis for mcl-s22, lot 110903mf4, found that the clevis that holds the jaws is broken.The tube is bent.There are impacts on the jaws.However, there is no missing part; there is no risk of fragment in the patient.No material or manufacturing defect was found.The most probable cause of this breakage is an excessive effort during the use or the reprocessing of the instrument.(b)(4).
|