Catalog Number 1012581-19
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/05/2015
Event Type  Injury  
Event Description
It was reported that the procedure was to treat a perforation in the saphenous vein graft (svg) to the left anterior descending (lad) artery.A 3.50x19mm graftmaster stent was advanced and deployed; however, the stent graft did not fully seal the perforation.A 3.5x16 mm rx graftmaster was then advanced and implanted to successfully seal the perforation.There was no adverse patient sequelae and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).It is indicated that the device is not returning for evaluation; therefore, a failure analysis of the complaint device could not be completed.A review of the lot history record revealed no non-conformances that would have contributed to this complaint.A query of the electronic complaint handling database revealed no other incidents for stent graft leak reported from this lot.Based on the information reviewed, there is no indication of a product deficiency.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Type of Device
CORONARY STENT GRAFT
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key4481329
MDR Text Key5346706
Report Number2024168-2015-00589
Device Sequence Number1
Product Code MAF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H000001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 01/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2015
Device Catalogue Number1012581-19
Device Lot Number4013141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/21/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age87 YR
Patient Weight85
-
-