• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM, INC. AFFINITY 4 BED; BIRTHING BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM, INC. AFFINITY 4 BED; BIRTHING BED Back to Search Results
Model Number 3700
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/06/2015
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found all four casters rubber wheels are worn out from wear and tear causing bed to slightly move.Per the hill-rom service manual, the affinity three birthing bed and affinity four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventative maintenance.Check the tires for cuts, wear, tread life, etc.Apply the brake, and check to ensure that the bed will not move.If the bed moves, inspect it for wear, and adjust if required.Apply the steering pedal and check the steering to ensure proper locking action when activated.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2012-2014.It is unk if the facility performed any other preventative maintenance on this bed.The technician replaced all four brake casters to resolve the issue.Based on this info, no further action is required.
 
Event Description
Hill-rom received a report from the account stating that the brakes were not holding.The bed was located at the account.There was no pt/user injury reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED
Type of Device
BIRTHING BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
john cummings
1069 state rte 46 e
batesville, IN 47006
8129312869
MDR Report Key4483463
MDR Text Key21214876
Report Number1824206-2015-00149
Device Sequence Number1
Product Code HDD
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Unknown
Type of Report Initial
Report Date 01/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/06/2015
Initial Date FDA Received01/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-