This freedom a/c power supply was not in patient use.The freedom a/c power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that the freedom a/c power supply had a broken plug and would not function properly.Freedom ac power supply was returned to syncardia for evaluation.Visual inspection of the freedom ac power supply revealed that the cable connector cover was installed backwards.This is the root cause of the customer-reported inability to connect the ac power supply to a freedom power adaptor.The ac power supply cable was disassembled, the connector key slot was aligned with the connector cover key tab and the correct mating configuration was obtained.After the ac power supply cable was reassembled, the ac power supply passed all testing requirements.This issue was identified to be isolated to one lot that was inspected at syncardia.In response to an unrelated issue regarding power supply connectors coming apart, the inspector performed a verification to ensure the connectors' interior threads were properly mated with adhesive.The verification required the inspector to remove the connector covers.Therefore, it is possible that the cable connector cover was installed incorrectly during reassembly at syncardia during incoming inspection.Incoming inspection requirements were improved to clearly identify the correct removal and replacement of the connector cover.This failure mode poses a low risk to a patient because the freedom ac power supply was not in use by a patient at the time of the customer-reported issue.In addition, it would not prevent a freedom driver from performing its life-sustaining functions.The freedom driver is equipped with redundant power sources, including multiple rechargeable freedom onboard batteries and a car charger.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
|