• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY CAROTID WALLSTENT®; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY CAROTID WALLSTENT®; STENT, CAROTID Back to Search Results
Model Number M001719010
Device Problems Device Damaged Prior to Use (2284); Activation, Positioning or Separation Problem (2906)
Patient Problem No Patient Involvement (2645)
Event Date 01/21/2015
Event Type  malfunction  
Event Description
It was reported that the stent came off the catheter.A 8x21,5f, 135cm carotid wallstent® was selected however the stent came off the catheter during preparation.The device never entered the patient¿s body.No patient complications were reported.
 
Manufacturer Narrative
(b)(4).The complaint device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred without direct patient contact.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT®
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4484819
MDR Text Key5502784
Report Number2134265-2015-00513
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/12/2018
Device Model NumberM001719010
Device Catalogue Number71-901
Device Lot Number17273096
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/22/2015
Initial Date FDA Received02/04/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/18/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-