On (b)(6) 2015, the patient (b)(6) underwent an implant procedure.During the procedure, the doctor was unable to remove the stylet from the lead intended for use.In turn, the stylet broke inside of the lead as a result of the doctor applying force.Another lead was used to successfully complete the procedure.The procedure was extended by an hour.Patient information is unknown.
|
Evaluation codes: results the complaint of ¿can¿t remove stylet¿ was confirmed.As received, the stylet returned had five abnormal bends 3cm, 16cm, 28cm, 35cm, and another 55cm away from the tip; and with a broken handle.Optical inspection of the returned lead body was clear.Both stimulation and terminal ends were in good condition, and no broken wires or kinks in the lead body were observed.The inner stylet lumen (stylet tubing) was buckled/collapsed approximately 16cm from the terminal end of the lead.A lab stylet was inserted and extracted with slight resistance.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|