• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN® GOLD-TITE® HEXED SCREW; ABUTMENT SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN® GOLD-TITE® HEXED SCREW; ABUTMENT SCREW Back to Search Results
Catalog Number IUNIHG
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem No Code Available (3191)
Event Date 01/20/2015
Event Type  Injury  
Event Description
The doctor states the abutment screw broke off in the implant and was not retrievable.The implant was removed and the site was grafted for future implantation.
 
Manufacturer Narrative
The investigator¿s visual inspection show the screw has fractured and remains stuck inside of the implant.A complete dimensional analysis cannot be performed due to the damage of the as returned product.The lot number was not provided for the abutment screw associated with this event, complaint history review did not provide any indication of a manufacturing deviation that would contribute to this event.A definitive root cause has not been determined.No results can be drawn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN® GOLD-TITE® HEXED SCREW
Type of Device
ABUTMENT SCREW
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
dania perez
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key4495264
MDR Text Key15105002
Report Number0001038806-2015-00072
Device Sequence Number1
Product Code NHA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PNHA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 01/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberIUNIHG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NIOSS411
Patient Outcome(s) Required Intervention;
Patient Age70 YR
-
-